Capillary
electrophoresis
From Wikipedia, the free encyclopedia
Revision as of 16:40, 11 February 2007 by FANSTARbot
(Talk | contribs)
Capillary
electrophoresis
(CE) can be used to separate ionic species
by their charge and frictional forces. In traditional
electrophoresis, electrically charged analytes move
in a conductive
liquid
medium under the influence of an electric field. Introduced in the 1960s, the technique
of capillary electrophoresis (CE) was designed to
separate species based on their size to charge ratio
in the interior of a small capillary filled
with an electrolyte.
Contents |
[edit] Instrumentation
The
instrumentation needed to perform capillary electrophoresis
is relatively simple. A basic schematic of a capillary electrophoresis system is shown in
figure 1. The system's main components are
a sample vial, source and destination vials, a capillary,
electrodes, a high-voltage power supply,
a detector, and a data output and handling device.
The source vial, destination vial and capillary are
filled with an electrolyte such as an aqueous buffer
solution. To introduce the sample, the capillary inlet
is placed into a vial containing the sample and then
returned to the source vial (sample is introduced
into the capillary via capillary
action, pressure, or siphoning). The migration
of the analytes is then initiated by an electric field
that is applied between the source and destination
vials and is supplied to the electrodes by the high-voltage
power supply. It is important to note that all ions,
positive or negative, are pulled through the capillary
in the same direction by electroosmotic flow, as will
be explained. The analytes separate as they migrate
due to their electrophoretic mobility, as will be
explained, and are detected near the outlet end of
the capillary. The output of the detector is sent
to a data output and handling device such as an integrator or computer. The
data is then displayed as an electropherogram, which
reports detector response as a function of time.
Separated chemical compounds appear as peaks with different
retention times in an electropherogram.[1]
[edit] Detection
Separation
by capillary electrophoresis can be detected by several
detection devices. The majority of commercial systems
use UV or UV-Vis absorbance as their primary mode of detection. In these systems,
a section of the capillary itself is used as the detection
cell. The use of on-tube detection enables detection
of separated analytes with no loss of resolution.
In general, capillaries used in capillary electrophoresis
are coated with a polymer
for increased stability. The portion of the capillary
used for UV detection, however, must be optically
transparent. Bare capillaries can break relatively
easily and, as a result, capillaries with transparent
coatings are available to increase the stability of
the cell window. The path length
of the detection cell in capillary electrophoresis
(~ 50 micrometers)
is far less than that of a traditional UV cell (~
1 cm). According
to the Beer-Lambert law, the sensitivity of the detector
is proportional to the path length of the cell. To
improve the sensitivity, the path length can be increased,
though this results in a loss of resolution. The capillary
tube itself can be expanded at the detection point,
creating a "bubble cell" with a longer path
length or additional tubing can be added at the detection
point as shown in figure 2. Both of these methods,
however, will decrease the resolution of the separation.[2]
Figure 2: Techniques for increasing the pathlength of the capillary: a.) a
bubble cell and b.) a z-cell (additional tubing).[1]
Fluorescence
detection can also be used in capillary electrophoresis
for samples that naturally fluoresce or are chemically
modified to contain fluorescent
tags. This mode of detection offers high sensitivity
and improved selectivity for these samples, but cannot
be utilized for samples that do not fluoresce. The
set-up for fluorescence detection in a capillary electrophoresis
system can be complicated. The method requires that
the light beam be focused on the capillary, which
can be difficult for many light sources.[2] Laser-induced fluorescence has been used in CE systems with detection
limits as low as 10-18 to 10-21
mol. The sensitivity of the technique is attributed
to the high intensity of
the incident
light and the ability to accurately focus the light
on the capillary.[1]
In
order to obtain the identity of sample components,
capillary electrophoresis can be directly coupled
with mass
spectrometers or Surface-Enhanced Raman Spectroscopy
(SERS). In most systems,
the capillary outlet is introduced into an ion source
that utilizes electrospray ionization (ESI). The resulting
ions are then analyzed by the mass spectrometer. This
set-up requires volatile
buffer solutions, which will affect the range of separation
modes that can be employed and the degree of resolution
that can be achieved.[2] The measurement and analysis are mostly done with
a specialized gel analysis software.
For
CE-SERS, capillary electrophoresis eluants can be
deposited onto a SERS-active substrate. Analyte retention
times can be translated into spatial distance by moving
the SERS-active substrate at a constant rate during
capillary electrophoresis. This allows the subsequent
spectroscopic technique to be applied to specific
eluants for identification with high sensitivity.
SERS-active substrates can be chosen that do not interfere
with the spectrum of the analytes.[3]
[edit] Modes of separation
The
separation of compounds by capillary electrophoresis
is dependent on the differential migration of analytes
in an applied electric field. The electrophoretic
migration velocity (up)
of an analyte toward the electrode of opposite charge
is:
up = μpE
where
μp is the electrophoretic mobility and
E is the electric field strength. The electrophoretic
mobility is proportional to the ionic charge of a
sample and inversely proportional to any frictional forces present in the
buffer. When two species in a sample have different
charges or experience different frictional forces,
they will separate from one another as they migrate
through a buffer solution. The frictional forces experienced
by an analyte ion depend on the viscosity (η) of the medium and the size and shape of the
ion.[2] Accordingly, the electrophoretic mobility of an
analyte at a given pH
is given by:
![]()
where
z is the net charge of the analyte and r is the Stokes radius of the analyte. The Stokes
radius is given by:
![]()
where
kB is the Boltzmann constant, and T
is the temperature,
D is the diffusion coefficient. These equations indicate
that the electrophoretic mobility of the analyte is
proportional to the charge of the analyte and inversely
proportional to its radius. The electrophoretic
mobility can be determined experimentally from the
migration time and the field strength:
![]()
where
L is the distance from the inlet to the detection
point, tr
is the time required for the analyte to reach the
detection point (migration time), V
is the applied voltage (field strength), and Lt
is the total length of the capillary.[2]
Since only charged ions are affected by the electric
field, neutral analytes are poorly separated by capillary
electrophoresis.
The
velocity of migration of an analyte in capillary electrophoresis
will also depend upon the rate of electroosmotic flow (EOF) of the buffer solution.
In a typical system, the electroosmotic flow is directed
toward the negatively charged cathode so that the buffer flows through the capillary from
the source vial to the destination vial. Separated
by differing electrophoretic mobilities, analytes
migrate toward the electrode of opposite charge.[1] As a result, negatively charged analytes are attracted
to the positively charged anode, counter to the EOF, while positively charged analytes are
attracted to the cathode, in agreement with the EOF as depicted in figure
3.
The
velocity of the electroosmotic flow, uo
can be written as:
uo = μoE
where
μo is the electroosmotic mobility, which
is defined as:
![]()
where
ζ is the zeta
potential of the capillary wall, and ε
is the relative permittivity of the buffer solution.
Experimentally, the electroosmotic mobility can be
determined by measuring the retention time of a neutral
analyte.[2]
The velocity (u)
of an analyte in an electric field can then be defined
as:
up + uo
= (μp + μo)E
Since
the electroosmotic flow of the buffer solution is
generally greater than that of the electrophoretic
flow of the analytes, all analytes are carried along
with the buffer solution toward the cathode. Even
small, triply charged anions can be redirected to
the cathode by the relatively powerful EOF of the
buffer solution. Negatively charged analytes are retained
longer in the capilliary due to their conflicting
electrophoretic mobilities.[1]
The order of migration seen by the detector is shown
in figure 3: small multiply charged cations
migrate quickly and small multiply charged anions are retained
strongly.[2]
Electroosmotic
flow is observed when an electric field is applied
to a solution in a capillary that has fixed charges
on its interior wall. Charge is accumulated on the
inner surface of a capillary when a buffer solution
is placed inside the capillary. In a fused-silica
capillary, silanol
(Si-OH) groups attached to the interior wall of the
capillary are ionized to negatively charged silanoate
(Si-O-) groups at pH values greater than
three. The ionization of the capillary wall can be
enhanced by first running a basic solution, such as
NaOH or KOH through the capillary prior to introducing
the buffer solution. Attracted to the negatively charged
silanoate groups, the positively charged cations of
the buffer solution will form two inner layers of
cations (called the diffuse double layer or the electrical
double layer) on the capillary wall as shown in figure
4. The first layer is referred to as the fixed
layer because it is held tightly to the silanoate
groups. The outer layer, called the mobile layer,
is farther from the silanoate groups. The mobile cation
layer is pulled in the direction of the negatively
charged cathode when an electric field is applied.
Since these cations are solvated, the
bulk buffer solution migrates with the mobile layer,
causing the electroosmotic flow of the buffer solution.
Other capillaries including Teflon capillaries
also exhibit electroosmotic flow. The EOF of these
capillaries is probably the result of adsorption
of the electrically charged ions of the buffer onto
the capillary walls.[1] The rate of EOF is dependent on the field strength
and the charge density of the capillary wall. The
wall's charge density is proportional to the pH of
the buffer solution. The electroosmotic flow will
increase with pH until all of the available silanols
lining the wall of the capillary are fully ionized.[2]
[edit] Efficiency and resolution
The
number of theoretical plates, or separation efficiency,
in capillary electrophoresis is given by:
![]()
where
N is the number of theoretical
plates, μ is the apparent
mobility in the separation medium and Dm
is the diffusion coefficient
of the analyte. According to this equation, the efficiency
of separation is only limited by diffusion and is
proportional to the strength of the electric field.
The efficiency of capillary electrophoresis separations
is typically much higher than the efficiency of other
separation techniques like HPLC. Unlike HPLC, in
capillary electrophoresis there is no mass transfer
between phases.[2] In addition, the
flow profile in EOF-driven systems is flat, rather
than the rounded laminar
flow profile characteristic of the pressure-driven
flow in chromatography columns as shown in figure
5. As a result, EOF does not significantly contribute
to band broadening as in pressure-driven chromatography.
Capillary electrophoresis separations can have several
hundred thousand theoretical plates.[4]
The
resolution (Rs)
of capillary electrophoresis separations can be written
as:

According
to this equation, maximum
resolution is reached when the electrophoretic and
electroosmotic mobilities are similar in magnitude and opposite in sign. In addition, it can be seen
that high resolution requires lower velocity and,
correspondingly, increased analysis time.[2]
[edit] Related techniques
As
discussed above, separations in a capillary electrophoresis
system are typically dependent on the analytes having
different electrophoretic mobilities. However, some
classes of analyte cannot be separated by this effect
because they are neutral (uncharged) or because they
may not differ significantly in electrophoretic mobility.
However, there are several techniques that can help
separate such analytes with a capillary electrophoresis
system. Adding a surfactant to the electrolyte can
facilitate the separation of neutral compounds by
micellar electrokinetic chromatography.
Charged polymers such as DNA can be separated by
filling the capillary with a gel matrix that retards
longer strands more than shorter strands. This is
called capillary gel
electrophoresis. This is a high-resolution
alternative to slab gel electrophoresis. Some capillary electrophoresis
systems can also be used for microscale liquid chromatography or capillary electrochromatography. A capillary electrophoresis
system can also be used for isotachophoresis
and isoelectric focussing.
[edit] References
- ^ a b c d e f
- ^ a b c d e f g h i j
- ^ Lin H.; Natan, M.; Keating, C. Anal. Chem.
2000, 72, 5348-5355.
- ^ Skoog, D.A.; Holler, F.J.; Nieman, T.A. "Principles
of Instrumental Analysis, 5th ed." Saunders
college Publishing: Philadelphia, 1998.
References
not cited in-line:
- Terabe, S.; Otsuka, K.; Ichikawa, K.;
Tsuchiya, A.; Ando, T. Anal. Chem. 1984,
56, 111.
- Terabe, S.; Otsuka, K.; Ichikawa, K.;
Tsuchiya, A.; Ando, T. Anal. Chem. 1984,
56, 113.
- Foley, J.P. Anal. Chem. 1990,
62, 1302.
- Carretero, A.S.; Cruces-Blanco, C.;
Ramirez, S.C.; Pancorbo, A.C.; Gutierrez, A.F. J.
Agric. Food. Chem. 2004, 52, 5791.
- Cavazza, A.; Corradini, C.; Lauria,
A.; Nicoletti, I. J. Agric. Food Chem. 2000,
48, 3324.
- Rodrigues, M.R.A.; Caramao, E.B.; Arce,
L.; Rios, A.; Valcarcel, M. J. Agric. Food Chem.
2002, 50, 425.
[edit] Companies
- Micronit Microfluidics - Glass chips (lab-on-a-chip) and
microfluidic tools for capillary electrophoresis
- Beckman Coulter [1] Instruments for Pharmaceutical Quality Control, CdT
analysis, small molecule analysis and protein characterisation.
- CE Resources [2] CE Instruments for biomedical and chemical analysis.
The CE-P2 is currently the only portable automated
CE system in the world.
[edit] See also
DNA Separation by Silica Adsorption
Retrieved from "http://en.wikipedia.org/wiki/Capillary_electrophoresis"




